ISO80369-7 Figure C.4 Male Reference Luer Lock Connector For Testing Female Luer Connectors Leakage

ISO80369-7 Figure C.4 Male Reference Luer Lock Connector For Testing Female Luer Connectors Leakage

  • Model:

ISO 80369-7:2017 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications

 

ISO80369-7 Figure C.4 Male Reference Luer Lock Connector For Testing Female Luer Connectors Leakage 0

a Maximum distance from tip of male luer lock connector to the first complete profile of the internal thread
(see Table B.3, dimension t).
b Double-start, right-hand thread of 2,5 mm pitch.
NOTE : R may be 0,5 mm (maximum) × 45° chamfer.

Figure C.4 — Male reference luer lock connector for testing female luer connectors for leakage, separation from unscrewing, stress cracking and non-interconnectable characteristics

ISO80369-7 Figure C.4 Male Reference Luer Lock Connector For Testing Female Luer Connectors Leakage 1

ISO80369-7 Figure C.4 Male Reference Luer Lock Connector For Testing Female Luer Connectors Leakage 2