iso 80369 test equipment

iso 80369 test equipment

  • Model:

ISO80369-20 Pressure generator for Positive air pressure,Subatmospheric-pressure air, positive pressure liquid

ISO80369 Testing Equipment,Luer Gauge,ISO80369 Gauge,ISO5356 GaugeISO80369 Testing Equipment,Luer Gauge,ISO80369 Gauge,ISO5356 Gauge



Purchase list:

1Luer gauge


according to ISO80369-7-Figure C.1 Female reference luer lock connector for testing male luer connectors for leakage, separation from unscrewing, stress cracking and non-interconnectable characteristics


2Luer gauge


according to ISO80369-7-Figure C.2 Male reference luer slip connector for testing female luer connectors for leakage, separation from axial load, stress cracking and non-interconnectable characteristics


3Luer gauge


according to ISO80369-7-Figure C.3 Female reference connector for testing male luer lock connector for separation from axial load and resistance to overriding


4Luer gauge


according to ISO80369-7-Figure C.4 Male reference luer lock connector for testing female luer connectors
for leakage, separation from unscrewing, stress cracking and non-interconnectable characteristics


5Luer gauge


according to ISO80369-7-Figure C.5 Female reference luer slip connector for testing male luer connectors for leakage, separation from axial load, stress cracking and non-interconnectable characteristics


6Luer gauge


according to ISO80369-7-Figure C.6 Male reference connector for testing female luer lock connector for separation from axial load and resistance to overriding


7CONNECTOR ASSEMBLY DEVICE for ISO 80369-20


The device is used to apply the required force and torque simultaneously when connecting a reference connector with a test sample. The device may also be used to apply a specified force-only or torque-only. Assembly Force : 26.5N - 27.5N Assembly Torque : 0.08 - 0.12Nm for lock connectors


8


pressure and positive-pressure liquid leakage tester


Control system Standard:ISO80369-20 Annex B and Annex C with mini printer.
9


Subatmospheric-pressure air leakage tester


Control system Standard:ISO80369-20 Annex D with mini printer.
10Pressure generator


Positive air pressure,Subatmospheric-pressure air, positive pressure liquid


11


Resistance to separation from axial load tester



Standard:ISO80369-20 Annex F Measuring range:0-500N resolution:0.01N Force accuracy:<±1% speed:10N/s Time:0-9999s with mini printer.