ZR-1210 Model Respirator Resistance Detector With High Definition LCD Liquid Crystal Display
ZR-1210 Model Respirator Resistance Detector With High Definition LCD Liquid Crystal Display
- Model:
ZR-1210 Model Respirator Resistance Detector With High Definition LCD Liquid Crystal Display
ZR-1210model respirator resistance detector
Product introduction
Zr-1210 respirator resistance tester is used to measure the inspiratory and expiratory resistance of the mask under specified conditions.It is applicable to the inspection and inspection of mask products by mask manufacturers and national labor protection equipment inspection institutions.
Executive standard
GB 2626-2019 Respiratory protective self-absorption filter respirator against particulate matter
Technical characteristics
High definition LCD liquid crystal display;
Electronic flowmeter, high flow control precision;
Automatic constant current control, automatic sample qualification;
Two modes of breath detection and breath detection can be set.
Large capacity data storage, real-time storage of detection data;
Can be exported through the U disk or thermal printer to print historical data;
Parameters such as pressure difference and sample number can be set;
Automatic protection for fault detection.
Technical indicators
main parameters | parameters | resolution | Maximum permissible error |
test flow | (85±1)L/min | 0.1L/min | ±1% |
Pressure range | ±1000Pa | 0.1Pa | ±1% |
Data storage | 500groups | ||
Working power | AC 220V±10% 50Hz | ||
Overall dimensions(W×D×H) | (335×280×255)mm | ||
Host weight | About11kg | ||
Host power consumption | <80W |